TBS 3000 HTHS Viscometer-环球(香港)科技有限公司
Product
Location:Home > Our Products > Applications > Industrial Lubrication > TBS 3000 HTHS Viscometer
Product information
TBS 3000 HTHS Viscometer
Category:Industrial Lubrication
Method Standard:ASTM D4683,D6616,IP 370, CEC L-36-A90
Application:Measurement of High Temperature High Shear Rate Rotational (Absolute) Viscosity for Engine Oils
Brand:TANNAS & King
Description Features Test Methods Instrument Specifications

High-Temperature, High Shear Rate Rotational (Absolute) Viscosity: The coaxial Rotor/Stator design permits the exact measurement of rotor position and the torque response of the liquid’s resistance to flow (viscous friction), which determines the apparent fluid viscosity. Applying a constant and linear shear rate profile continually to the fluid makes the TBS an absolute viscometer where the ‘true’ shear rate is calculated from known dimensions and speed of rotation. The TBS measures the viscosities of fresh and used oils at multiple select shear rates and temperatures. This proves particularly useful with multi-grade oils and their influence on fuel efficiency.

Developed in 1979 and under several patents, the TBS Viscometer became the world’s first high shear viscometer. It remains the modern benchmark ‘referee’ instrument for HTHS viscometry due to its innovative measurement technique, robust design, and notable operational versatility.

This newest TBS model is completely redesigned and re-innovated providing multiple enhancements to electrical and temperature components, automation, calibrations, and oil injection features, dramatically improving the overall user experience. These advancements promote simplified operations and less reference oil usage with no separate components while offering the same or better precision and robustness as previous TBS models.

 


· Integrated Custom Control System with custom PCB Boards and built-in TBS Touch Software – Eliminates computer, DAQ Boards, and associated cabling.

· Improved operating stability to varying lab environment conditions.

· Built-in 40-position AutoSampler with auto single-injection mechanism.

· Incorporates (10) easily replaceable Reference Oil containers (Quart, 1/2 Gal. or Gal.).

· Custom LED lighting scheme based on operating condition.

· Innovative temperature control system – no external cooling bath for 80° operation.

· Auto-monitoring of Waste Oil container with Overfill Alarm.

· Remote data access with Ethernet connection.

· Barcode reader compatible.


· ASTM D4683 | D6616

· CEC L-36-90

· IP 370

· JPI-5S-36-03

· Reference Technique for ASTM D4741 and D5481


Dimensions (W x D x H)

Benchtop: 66 x 66 x 48 cm (26 x 26 x 19 inches)

Weight

~ 38.5 kg | (85 lbs.)

Voltage

100-240 VAC @ up to 6 Amp

Frequency

50/60 Hz.

Viscosity Range

1 to 10-15 mPa•s (cP) depending on temperature and shear rate

Sample Volume

Recommended 40 mL for chase-flush technique

Spindle Speed

Up to 8,000 RPM

Sample Test Time

Approximately 12 minutes per sample

AutoSampler Capacity

Auto-inject 1 to 40 sample sequentially

Operating Temperatures

Constant Temperature Control: 80°C, 100°C, and 150°C (±0.1°C)

Shear Rates

500,000 sec-1 to (3+) million sec-1 depending on viscosity and temperature

Read-out

Digital Touchscreen Display of Temperature, Torque and other operating conditions.

Safety

Safe restart (rotor motor and stator heater off) in case of power loss, CE Mark