Features & Benefits
IQT®-XLM Software
The main system operating screen clearly displays all operating parameters, using large, color-coded bar graphs to indicate to the user whether a parameter is within allowable limits. The bar graphs are sized so that a basic check of the instrument’s condition and test progress can be done from a distance.
New intuitive and convenient user interface with:
· Simplified access to calibration and diagnostic functions
· Advanced diagnostics for condition-based maintenance
· Multiple language capabilities
Features
Touch screen HMI with industrial PC
Windows® based user interface
Electrical Connection: 20A,208-240 VAC, 50/60Hz
Integral cooling system with built-in Uninterruptible Power Supply (UPS): > 3.5 hrs of run time for the cooling system and computer
Low Sample Volume: 35 ml for cleaning, 30 ml for test run
Measurement Results
Ignition delay (ms) and Derived Cetane Number (DCN), Correlation of DCN and Cetane Number (CN) recognized by ASTM (versus D613) and CEN (versus ISO 5165)
Repeatability
0.53 @ 40 DCN
0.65 @ 50 DCN
0.89 @ 70 DCN (ASTM D6890-21)
Derived Cetane Number Range
31.5 to 75.1DCN (ASTM D6890 21 scope)
30.4 to 78.9 DCN (applicability of ASTM D6890 precision functions per ASTM D6300)
Compressed Gas Requirements
Compressed air at 12.1 bar (175 psi)
Compressed air (19.9-21.9% O2) at 21.4 bar (310 psi)
Compressed nitrogen at 3.5 bar (50 psi)
Reference Materials
n-heptane 99.5%, octane engine grade (calibration reference material, check standard)
Methylcyclohexane 99% (verification material)
Reference diesel fuel (quality control sample)